Draft - 1 Aug 08
Technical Alarms (INOPs)
5-10
Alarms
Technical Alarms (INOPs)
Technical Alarms, or INOPs (inoperative conditions), are sourced at the
transceiver, the ST/AR algorithm running at the Information Center, or TeleMon
Companion Monitor. They identify inoperative conditions (that is conditions
where the system is not operating properly and therefore cannot measure or
detect alarm conditions reliably). There are four levels of Technical Alarms:
•
Severe
- Monitoring and alarms disabled. Audible tone at the Information
Center. Must be acknowledged by a clinician.
•
Hard
- Monitoring and alarms are disabled. Audible tone at the
Information Center.
If the hard INOP is "latched", the sound will be silenced, but the message
will remain on the display until resolution of the offending condition.
•
Soft
- Monitoring and alarms remain active. No audible tones are
generated.
•
Red/Yellow
- Replace Battery and ECG Leads Off INOPs may be
configured to display as either Red or Yellow Technical Alarms.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...