Telemetry Functions at the Information Center
9-14
Telemetry Functions at the Information Center & TeleMon
SpO
2
Alarm
Turn SpO
2
alarms on/off at the
Information Center.
enable
(on)
disable
(off)
enable
SpO
2
Limits
High
Increment/decrement SpO
2
high
alarm limit by 1 (in %).
Limit maximum is 100. Limit
minimum is 51 (adult) or 31
(pediatric). High and low
limit must be at least 1%
apart.
100 (adult,
pediatric)
SpO
2
Limits
Low
Increment/decrement SpO
2
low
alarm limit by 1 (in %).
Limit maximum is 99. Limit
minimum is 50 (adult) or 30
(pediatric). High and low
limit must be at least 1%
apart.
90 (adult,
pediatric)
Unit Settings - Telemetry Setup
Control
Function
Settings
Factory
Default
Unit Settings - Default Leads
Control
Function
Settings
Factory
Default
Leadset
Allowed
Set the unit default leadset type.
All, EASI, No EASI, 3 wire
All
3-wire
Set the unit default lead.
I
,
II
,
III
II
5-wire, ECG1
Set the unit default lead.
I
,
II
,
III
,
MCL
,
AVR
,
AVL
,
AVF
,
V
II
5-wire, ECG2
Set the unit default lead.
I
,
II
,
III
,
MCL
,
AVR
,
AVL
,
AVF
,
V
V
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...