Audible Tones
Transceiver Controls
3-13
and pulse detection tone. The sounds can be turned off (muted) on a per patient
basis in the Telemetry Setup window.
Service
Sounds
Two additional sounds that you may hear occur during the labelling of
transceiver devices within the IntelliVue Clinical Network. The volume of these
sounds is not adjustable, and cannot be turned off. See “Label Assignment for
Replacement Transceiver” on page 11-2.
Auditory
Signal
(sound)
Function
How Used
Sound 1
(beep)
Spot Check
Announces a successful SpO
2
Spot Check
measurement.
Sound 1
(beep)
Check
Button
Confirms contact with the Information Center.
Sound 2
(low pitch
tone)
Pulse
Detection
Tone
Indicates pulse during SpO
2
Spot Check
measurement.
Service Sound
Description
beep beep repeated every
3 seconds, and all visual
indicators flashing
Unlabeled Device.
Equipment Label needs to be assigned at
Information Center. No monitoring.
Fast beep beep
and
alternate Leads Off
indicators flashing
Unacknowledged Labeled Device.
Equipment label has been assigned at
Information Center; press the Check button to
complete label assignment. No monitoring.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...