Draft - 1 Aug 08
Pairing Non-networked Devices
Pairing Monitoring Devices
10-23
Replace BatteryT INOP
displayed at the monitor
and the Information Center.
When operating wirelessly, the
bedside monitor is no longer
providing power to the
transmitter, and battery capacity
is now depleted. There is
no
monitoring
occurring at either
the monitor or the Information
Center.
Insert new AA batteries in
the transceiver.
Absence of successful
association sound/visual
indicator.
The bedside monitor and the
transceiver are not
communicating with each other.
Monitoring at the Information
Center only.
Repeat the device
assignment procedure.
Central: Tele Only INOP
When the transceiver is wirelessly
paired with an X2 patient monitor
(no label) docked with a larger
networked MP series monitor,
and the network connection is
lost.
Restore the monitor’s
network connection.
Check Pairing INOP
An X2 patient monitor with a
telemetry label is paired with a
larger networked MP series
monitor and subsequently the
transceiver is paired with the
same monitor. Only one
transceiver can pair with the
monitor.
Select the correct device to
be paired.
Condition
Description
Action
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...