Draft - 1 Aug 08
Pairing Non-networked Devices
Pairing Monitoring Devices
10-21
More Bed
Alarms
The More Bed Alarms INOP is displayed at the Information Center when a
physiological alarm or INOP is generated by the bedside that is not included in
the group of alarms that is transmitted to the Information Center by the
transceiver. The INOP will display using the corresponding severity color of the
actual alarm. For example, should the bedside monitor generate a "***DESAT
alarm", it will display at the Information Center as a red, !!!More Bed Alarms
INOP.
Warning
Warning
***DESAT and ***<Pressure> HIGH alarms from the bedside monitor will
appear as red, !!!More Bed Alarms INOPs at the Information Center.