AAnalyst 200 User’s Guide
3-19
Flash arrestor
PerkinElmer recommends the use of the Flash Arrestor (P/N N930-0068). The flash arrestor
provides a positive gas shutoff by checking the reverse flow under all conditions. It also stops the
gas supply when a flashback occurs. The flash arrestor can be directly connected to the acetylene
fuel line.
Piping/tubing/fittings
Use galvanized iron, steel, or wrought iron tubing. Never use copper tubing with acetylene. Joints
may be welded or made up of threaded or flanged fittings, typically stainless steel or aluminum,
or brass composed of less than 65% copper. Rolled, forged, or cast steel or malleable iron fittings
may also be used. Cast iron fittings cannot be used safely for acetylene lines.
Safe use of acetylene
•
Periodically check for the presence of acetylene in the laboratory atmosphere, especially
near the ceiling.
•
Never allow the outlet gauge pressure to exceed 103 kPa (1.03 bar, 15 psig); acetylene can
explode spontaneously above this pressure.
•
Prevent acetylene from coming into contact with copper, silver, mercury or gaseous
chlorine.
•
Solvent carryover.
If the acetylene cylinder pressure falls to below 600 kPa / 6.0 bar / 87 psig (at 20 °C), some of the
solvent used to stabilize the acetylene (such as acetone) may be carried over into the burner. This
could influence the characteristics of the burner with a resulting influence on the analytical
results. Change to a new acetylene cylinder when the cylinder pressure falls below this value.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......