Analyzing Samples
8-16
Analyzing Samples Using Linear Calibration
The following example describes how to run samples using the linear calibration technique. It
should be noted that the same steps are carried out when using the nonlinear calibration
technique. The only difference is in setting up the calibration type. Instead of selecting nonlinear
for the calibration type, you select nonlinear.
EXAMPLE:
Determining copper using linear calibration
The procedures in this example assume that you have performed the set-up procedures in
Chapters 3, 4, 5, and 6, in which case the instrument will be on and the flame ignited.
This example runs a blank, 3 standards, and 3 test samples. We instructed the autosampler to
perform a reslope after the second test sample.
1.
Carry out the following steps to duplicate our example:
2.
Prepare solutions. Set up the autosampler
3.
Load the autosampler
4.
Enter parameters
5.
Start the run
Detailed procedures follow.
Prepare solutions
In this example, we used a copper lamp and solutions of copper for our determinations. However,
you do not have to use copper. For your convenience in making substitutions we have included a
table of recommended concentrations for Ag, Mg, Mn, and Pb.
1.
Prepare about 200-250 mL of 1% v/v HNO
3
to use as a blank solution.
2.
Prepare about 200-250 mL each of three standards in 1% v/v HNO
3
of the element you want
to determine. Suggested standard concentrations for five elements are shown in the following
table:
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......