Maintenance
9-42
Maintaining Optimum Nebulizer Performance
To maintain optimum nebulizer performance, you must periodically clean the capillary
assembly, and occasionally replace the capillary (sample) tubing. See
Figure 9- 13 for an exploded view of the nebulizer.
Also, the nebulizer is only a part of the overall burner system. For optimum burner-system
performance, it is important to periodically clean the nebulizer, burner chamber, burner heads,
and the end cap.
When aspirating concentrated samples with complex matrices, it is especially important to keep
the burner chamber and nebulizer free of sample deposits. In addition to regular cleaning, it is
beneficial to periodically aspirate a dilute surfactant solution, such as 0.1% Triton X100, for
several minutes to keep the system clean.
The surfactant solution will encourage constant drainage, which is required for obtaining
optimum burner system performance.
When using the AAnalyst nebulizer, the impact bead may have to be replaced.
Also, to avoid
damage to the impact bead, the bead should be removed when cleaning the capillary holder
assembly. Steps for removing and replacing the bead are included in this instruction document.
Cleaning the nebulizer
1.
After you have taken the nebulizer apart, clean the components as follows:
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......