Operation of the Instrument
6-36
7.
Ignite the flame:
Touch the
On
side of the
Flame On/Off
switch.
The indicator in the switch lights up.
8.
The correct procedure to extinguish the flame is described in this chapter.
9.
If the system is set up correctly the flame is ignited using default values for the gas flows. If
you selected nitrous oxide as the flame type, the flame is ignited with an air-acetylene
mixture, with automatic changeover to nitrous oxide after several seconds.
NOTE
: It is recommended to warm up the burner head with air-acetylene flame for 15 minutes before
switching to nitrous oxide flame
10.
The flame should appear uniform and stable throughout its length. If it is not, even after
warming up, you may need to clean the burner head. See the Maintenance chapter in this
User’s Guide Manual.
The nitrous oxide-acetylene flame may split and form a
V
along the burner slot when it is first
ignited. As the burner head warms up, the
V
will close. This may take several minutes.
WARNING
When using nitrous-oxide, the flame should have a red plume. Increase the
oxidant flow until the red plume is 1½ inches to 1¾ inches high.
Select an air-acetylene flame:
•
If you are preadjusting the nebulizer.
•
If you have installed a different nebulizer, cleaned the nebulizer, or if the settings on the
nebulizer have been altered since it was last used.
•
If you are using the flame to heat the quartz tube atomizer of an optional MHS-15 Mercury
Hydride System.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......