![PerkinElmer aanalyst 200 User Manual Download Page 61](http://html1.mh-extra.com/html/perkinelmer/aanalyst-200/aanalyst-200_user-manual_1530191061.webp)
AAnalyst 200 User’s Guide
3-17
CAUTION
Risk of damage to the instrument
Water in the air supply can cause damage to the pneumatics.
Use only dry air. Bleed compressor tanks to remove water at least once a week
Air compressors are generally uncomfortably noisy to have in the immediate vicinity of the
instrument. Whenever possible, locate them at some distance from the laboratory in an area
providing suitable ventilation.
Air cylinders
Cylinders of compressed air can also be used but are recommended only as an emergency or
short-term solution. A standard #1 size air cylinder will last only about five hours. It is common
to find other than 20% oxygen in air cylinders. This can cause erratic burner operation and non-
reproducible analytical results and, in extreme cases, may cause a potential safety hazard.
In general, if cylinder air is to be used, it is important to specify compressed air rather than
breathing air and that the oxygen content is less than 20%.
WARNING
Warning: Flashback Hazard
PerkinElmer burner systems are designed for use with compressed air. The use
of oxygen can cause an explosion in the burner system, and oxygen-enriched air
can cause a flashback of the flame.
Never use oxygen or oxygen-enriched air with PerkinElmer burner systems.
Use only compressed air
.
Nitrous oxide supply requirements
WARNING
Warning: Nitrous oxide is a strong oxidant and an asphyxiant
Nitrous oxide is an asphyxiant; it can be considered a potential hazard
whenever it is stored or used in a confined space.
Spontaneous combustion may occur if nitrous oxide comes into contact with
grease, oil, or other organic material.
•
Store nitrous oxide cylinders in a well-ventilated area outside the
laboratory.
•
Do not store nitrous oxide cylinders in close proximity to flammable gases
(e.g., acetylene).
•
Make sure tubing and fittings carrying nitrous oxide are free of grease, oil,
or other organic material.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......