Installation
5-20
Install the Burner Assembly
1.
Carefully remove the foam packing piece from around the burner system.
2.
Guide the burner assembly into the tracks of the atomizer compartment.
3.
Push the burner assembly in so that the rear gas fittings line up with the gas connectors at the back of
the atomizer compartment. Push the burner assembly fully in until it clicks into place and pull the
lever forward to lock the burner assembly in place.
Connecting the Drain Vessel and the Drain Interlock
1.
Locate the drain vessel. The drain tube and interlock cable will be attached to the drain vessel
retainer cap.
2.
Place the drain vessel in
full view
on the floor in front of the instrument.
Never place the drain
vessel inside a
cabinet.
3.
Before connecting the drain tube to the burner assembly, pour about 250 mL of water into the free
end of the drain tube to fill the drain loop and to activate the drain float.
4.
Place the hose clamp over the free end of the drain tubing.
5.
Carefully install the free end of the drain tubing over the drain outlet, making sure that the inner
lining of the drain tube does not get damaged.
6.
Use a flat-blade screwdriver to tighten the hose clamp.
7.
Plug the white drain interlock connector into the drain interlock plug on the burner assembly.
Screen 5- 7 Drain hose and drain interlock
Further procedures for setting up the burner system are described in Chapter 6
, Operation of the
Instrument.
Drain Interlock
Connector
Drain Hose
Clamp
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......