AAnalyst 200 User’s Guide
10-13
Troubleshooting Deuterium Background Corrector
Problems
Background correction requires the use of both the primary and the continuum source. It is
important that these sources be of similar intensity and pass through exactly the same area of the
sample cell. There are three possible problems that may be observed when the background
corrector source lamp is not aligned or needs replacing:
•
Measurements are excessively noisy.
•
The intensity of the background source may be much less than the intensity of the primary
source.
•
Background absorbance may not be properly corrected. You can check this by measuring a
sample containing only matrix components with no analyte.
These problems indicate that the background corrector source is either not properly aligned or
should be replaced. Refer to the section called
Replacing the Deuterium Lamp
in
Maintenance
chapter if it needs replacing.
Possible Cause of Problem Check
and/or
Remedy
Background source
intensity is much
lower than the
primary source
May not need to use the
background corrector.
Determine if the background corrector
should be used. It is useful up to about
400 nm.
The intensity is not matched.
1). Reduce the hollow cathode or the
EDL current in order to reduce the
primary source intensity. This may
increase the noise for the
determination.
2). In addition (or alternatively)
increase the slit width (this may
increase the possibility of spectral
interferences or increase the
calibration curvature).
The deuterium lamp may need
replacing (B315-0160).
Replace the lamp.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......