10-9
Flame does not ignite
(continued)
The flame may not ignite if the ignitor
tube is not positioned directly above
the burner slot.
Try igniting the flame a second
or third time.
Burner flashes
back
Flashbacks can be caused by any of
the reasons listed below.
Carefully inspect the burner
system for possible causes,
referring to the ones listed
below. Reignite the flame only
after the problem has been
corrected. Contact a
PerkinElmer service
representative for assistance.
Fuel and oxidant gas lines may have
been inadvertently switched.
Check the connections at the
burner and at the gas supplies.
The burner system may leak.
Check the condition of all
o-rings (burner head, end cap,
o-ring on nebulizer front end
cap). Replace o-rings that show
signs of wear (cracking,
swelling). Inspect the inner
surface of the burner chamber
for pitting and replace if
necessary. Check drain
connections for leaks.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......