AAnalyst 200 User’s Guide
6-39
Optimizing the Burner System
Summary of the optimization procedure
1.
Perform the preliminary steps.
2.
Prepare the required solutions.
3.
Install a lamp for the analyte element.
4.
Start up the spectrometer.
5.
Set up the burner system.
6.
Perform the safety checks. Ignite the flame.
7.
Optimize the burner position. Optimize the nebulizer.
8.
Optimize the gas flow rates.
Preliminary steps
Before you optimize the burner system for your particular analyses, perform the following
preliminary steps:
1.
Set up the burner system and make sure that the correct burner head is installed.
2.
Check that the burner height has been correctly set.
3.
Make sure that you have preadjusted the nebulizer.
Preparing the solutions
The blank and standard solutions should be the same as those you will use when analyzing
samples.
1.
Prepare a blank solution. Use the same solvent as you use to prepare the standard solution.
2.
Prepare a standard solution of the analyte element.
The concentration of the standard solution should be in the linear range and ideally should be
close to the ‘sensitivity check’ concentration quoted in the
Recommended Conditions
(i.e.
give an absorbance reading of approximately A = 0.2).
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......