AAnalyst 200 User’s Guide
6-43
Rinsing the Burner System
Before you extinguish the flame or begin maintenance or disassembly procedures, you must rinse
the burner system. After you extinguish the flame, allow the burner head to cool.
To rinse the system
The procedure for rinsing the burner system depends on the type of samples that you have
analyzed.
Recommendations:
•
If the sample solutions contained organic solvents, use a multi-rinse procedure to remove all
traces of the solvents from the system.
•
If the sample solutions contained toxic substances, you must rinse all traces of these
substances out of the system.
•
If the sample solutions contained high concentrations of copper, silver or mercury salts,
which can form unstable acetylides, you must rinse all traces of these substances out of the
system.
Procedures
If you used only aqueous solutions during the analyses.
•
Aspirate deionized water for 5 minutes.
If you used organic solvents during the analyses.
•
Aspirate for 5 minutes an organic solvent that is miscible with the samples that have just
been aspirated.
•
Aspirate for 5 minutes methanol or another solvent that is miscible both with water and the
solvent used previously.
•
Aspirate 1% (v/v) nitric acid solution for 5 minutes.
•
Aspirate deionized water for 5 minutes.
To extinguish the flame
1.
With the flame still burning, aspirate the correct rinsing solutions to rinse the burner system.
Either aspirate the solutions manually or use the flame autosampler.
2.
Touch the
Flame
tab to display the Flame page.
3.
Touch the
Off
side of the Flame
On/Off
switch. The indicator in the switch extinguishes.
The system uses a predetermined shutdown sequence to extinguish the flame safely.
4.
Shut down the gases to the spectrometer at source.
5.
Touch the
Bleed Gases
button
to depressurize the gas lines.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......