AAnalyst 200 User’s Guide
9-9
Burner Maintenance Checklist
Maintenance checklists for the burner system are presented below.
Burner system, maintenance checklist
If you analyze samples that have ‘dirty’ matrices, such as blood, sugar or oil, you will probably have to
clean the components more often than if you were analyzing ‘clean’ samples, such as drinking water.
Burner system
Always rinse the burner system correctly before extinguishing the flame.
If you are analyzing
samples with high solids content
: Clean the burner chamber at
least once daily.
If you are using
organic solvents
: Clean the burner chamber before you switch to
aqueous solutions. Empty the drain vessel and dispose of hazardous solutions
properly.
If you are analyzing
samples containing high concentrations of copper, silver, or
mercury
: Thoroughly flush the burner chamber and drain tube immediately after
completing such analyses. Empty the drain vessel and dispose of hazardous solutions
properly.
Burner head
Inspect the burner head for deposits before igniting the flame. The burner should
provide an even flame over the length of the burner slot. An uneven flame may
indicate that the slot needs cleaning. See page 9-19 for cleaning procedure.
Check the width of the burner slot periodically.
Nebulizer
If there is an obstruction in the nebulizer, clear it using a copper cleaning wire.
Drain system
Flush the drain system thoroughly at the end of each working day by pouring water
down the neck of the burner chamber.
Drain vessel
Empty the drain vessel regularly.
Do not
allow it to fill with corrosive, toxic, or
organic solutions. Dispose of hazardous solutions properly.
Drain trap
Periodically remove the drain trap and clean.
Drain tube
Inspect the drain tube regularly for wear and/or chemical attack. Replace it if it has
deteriorated.
O-rings
Check the condition of the o-rings regularly: burner end cap o-ring; o-ring in neck of
burner chamber; three o-rings on the nebulizer.
Gas hoses
Check the gas hoses regularly for leaks, particularly at the fittings, using a proprietary
leak testing fluid. Call a PerkinElmer service engineer if a leak is detected.
Acetylene cylinder If the cylinder pressure falls to below 600kPa (6bar; 85 psig) some of the solvent
used to stabilize the acetylene (e.g. acetone) may be carried over into the burner. This
could influence the characteristics of the burner. Always change to a new acetylene
cylinder when the cylinder pressure falls to below this value.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......