![PerkinElmer aanalyst 200 User Manual Download Page 171](http://html1.mh-extra.com/html/perkinelmer/aanalyst-200/aanalyst-200_user-manual_1530191171.webp)
AAnalyst 200 User’s Guide
7-13
Buttons on Flame page
Interlocks
The burner system has a safety feature that monitors gas pressure and checks for proper setup of
the burner head, nebulizer and drain system. A green check mark shows that all safety interlocks
are satisfied and that you can ignite the flame. A red cross mark indicates an unsatisfied safety
interlock. You must correct the hardware condition causing the problem before attempting to
ignite the flame. Touch the interlock field box to view the status of safety interlocks.
Screen 7- 8 Interlock Dialog
NOTE: If air-acetylene is used in place of N
2
0, the
N
2
0 pressure interlock will be ignored and flame
will light.
Flame On/Off
Touch the
On
side of this ‘switch’ to ignite the flame.
Touch the
Off
side of this ‘switch’ to extinguish the
flame.
In an emergency, press the
Red Emergency
button
located on the upper left side of the instrument. The
system will extinguish the flame.
Bleed Gases
After extinguishing the flame and shutting down the
gases at source, touch this button to depressurize the
gas lines.
Auto Scale Graph
Touch this button to resize the graph of the most recent
or current absorbance reading.
Auto Zero Graph
Touch this button to set the reading for the current
signal to zero on the graph.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......