Maintenance
9-44
Reassembly and Replacement
Reassembling the nebulizer
NOTE: If not already done, inspect the tiny o-rings on the capillary assembly for damage and replace
them if necessary. Lightly lubricate the surface of the o-rings with a minimum of Apiezon grease.
Make sure that no grease gets onto the tip of the capillary.
CAUTION
Do not add an excess of Apiezon grease. Lightly lubricate the surface of the
o-rings. Too much grease on the o-rings will affect instrument performance.
1.
Slide the spring back onto the capillary assembly (if it was removed).
2.
Slide the assembled capillary assembly into the nebulizer body.
3.
Line up the alignment key with the keyway in the nebulizer body.
4.
Push the capillary assembly in until it bottoms out in the nebulizer body.
5.
If you inserted the capillary assembly correctly, it should move freely in and out of the
nebulizer body. If it does not, remove and reinstall, paying particular attention to the
alignment key.
6.
Screw the adjusting nut onto the nebulizer body.
7.
Where applicable, refit the
fragile
impact bead
and spacer.
8.
Reattach the nebulizer sample tube to the nebulizer by screwing the connector into the
nebulizer body.
9.
Reinstall the nebulizer in the end cap.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......