AAnalyst 200 User’s Guide
7-45
Screen 7- 33 Command Prompt Window
Application Launch Failure
Use the following procedure below if the application fail to launch.
1.
Follow steps 1-4 from the
Connection Setup
procedure.
2.
Connect the keyboard to the E-Box.
3.
Power on the AA200 instrument and wait until the screen shows an arrow on top of the splash
screen.
4.
Timing is crucial: just as the arrow appears, use the keyboard to press the
ESC
key
continuously until a blue PerkinElmer Instrument
screen appears on the touch screen.
NOTE
: Press CTRL + ESC key to display the Start menu.
5.
Use the arrow keys on the keyboard to Select
Run
or touch
Run
on the touch screen.
6.
Enter the word ‘
killer.exe’
in the dialog box and touch
OK
.
NOTE
: Allow 10 seconds for Killer.exe to terminate all running windows and programs.
7.
Use the arrow keys on the keyboard to Select
Run
or touch
Run
on the touch screen.
8.
Enter the word
‘repllog’
in the dialog box and touch
OK
.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......