3.
Slide the lamp carefully into the lamp holder until it butts up against the end stop inside the holder.
4.
Connect the lamp cable to an EDL 1 port on the E-Box. Align the lamp.
Figure 6- 12 Assembling an EDL
Installing EDLs in the turret version
NOTE: EDLs may only be installed in positions 1 and 2 of the turret. Refer to Figure 6-15 for the assigned lamp
numbers on the turret (mount).
1.
Open the lamp door.
2.
Insert the coding plug over the receptacle on the lamp holder.
NOTE: Do not use coding plug for uncoded EDLs.
3.
Slide the lamp carefully into the lamp holder until it butts up against the end stop inside the holder.
4.
Depending on the lamp position of the turret, connect the lamp cable to the corresponding EDL port
on the E-Box. Align the lamp.
NOTE: Example: If the EDL lamp is installed in position 1 of the turret, connect the EDL cable to the EDL 1 port
on the E-Box.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......