Preparing Your Lab
3-14
In many countries, PerkinElmer offers an accessory blower and vent kit that will fulfill the
exhaust requirements for all atomic spectroscopy instruments, or your local PerkinElmer office
can supply you with the addresses of suitable suppliers.
In many countries electrical codes do not permit PerkinElmer service engineers to install the
blower and vent assembly.
Blower and vent kit
The blower and vent kit (P/N 0303-0406) available from PerkinElmer contains a rotary blower
with associated hardware, a hood to be located above the instrument, an adapter to connect the
blower and hood to suitable metal ducting, and installation instructions. The blower requires a
line voltage of 120 V or 230 V. The adapter and hood are made of stainless steel. The metal
ducting required to connect the hood to the blower and for the exhaust from the blower is not
included in the kit.
10 cm
(4 in.)
15 cm
(6 in.)
30 cm
(12 in.)
11 cm I.D.
(4.3 in.)
Figure 3- 4 Dimensions of a typical exhaust hoodGas Supply Requirements
The permanent installation of gas supplies is the responsibility of the user and should conform to
local safety and building regulations. PerkinElmer recommends that cylinders be stored outside
the laboratory and that gases be piped to the spectrometer system in approved gas lines.
•
The user must provide the gas supplies, regulators, connectors, and valves.
An air compressor is recommended for air. A double-stage or heated regulator is
recommended for nitrous oxide. Either the gas supplier or PerkinElmer can provide the
correct regulators and valves.
•
You must be able to reach the on/off valves easily and see the pressure indicators.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......