AAnalyst 200 User’s Guide
8-23
Running Samples in Emission Mode
The AAnalyst 200 also performs flame emission measurements at a fixed wavelength. This
option is useful when you do not have a lamp for the element you wish to determine or if you are
determining elements such as Na, K, Ca, or Ba, which are sensitively determined by emission.
A
setup solution
containing the element being determined must be aspirated while the
spectrometer wavelength is peaked and the high voltage (gain) on the photomultiplier is set. For
each element, the concentration of the setup solution should be the same as the concentration of
the highest calibration standard. Aspirating the setup solution produces an emission signal that is
required to peak the wavelength for the maximum signal and to set the photomultiplier voltage to
the proper value.
Setting Up for Emission
Refer to the section on Standard Conditions in
Analytical Methods for Atomic Absorption
Spectrometry
(Part Number 0303-0152) to find the recommended settings for wavelength, slit
width and flame type. Use these values to make appropriate settings.
1.
Switch on the AA200 instrument. See Chapter 6,
Operation of the Instrument
.
2.
The following page is the first page displayed once you start up the system. Select the
Flame
Technique in the Startup screen. Touch the
Details
bar on the screen to get details on the
Power-on Diagnostics. In the Details dialog all entries should be in green to indicate that all
systems are OK. If all the entries are green touch
OK
in the Startup screen.
3.
Turn on the Flame.
4.
If you want to recall an emission method, then do so at this time. When prompted, aspirate
the highest standard of the element being determined. This setup solution must be aspirated
while the spectrometer wavelength is peaked and the high voltage (gain) on the
photomultiplier is set.
If you do not have an emission method to recall, then proceed to step 5.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......