Maintenance
9-52
Replacing the Deuterium Arc Background Corrector
Lamp
WARNING
Since the instrument cover will be removed, this procedure should be done by a
skilled person who is trained in the servicing of optical and electrical
components and is aware of the hazards of high voltage electronics.
Equipment Needed
You’ll need to gather the following items in order to replace the deuterium lamp:
•
the deuterium arc lamp - P/N B315-0160
•
ultraviolet-absorbing safety glasses
•
flat bladed screwdriver
•
phillips head screwdriver
Removing the Deuterium Lamp Assembly
1.
Turn OFF the power switch and unplug the power cord from the E-Box.
WARNING
Wait a sufficient amount of time for the background corrector lamp to cool
before replacing it.
The lamp is very hot after use.
2.
Swing open the lamp compartment door.
3.
Locate the deuterium lamp assembly in the instrument (it is inside the large rectangular
block to the right of the lamp compartment).
4.
Unscrew the 4 corner phillips screws that hold the deuterium lamp plate in place. Don’t
let any of the 4 screws drop into the instrument.
5.
Carefully lift the assembly away from the housing and unplug the deuterium connector.
6.
Remove the assembly completely from the instrument and place it on the lab bench.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......