![PerkinElmer aanalyst 200 User Manual Download Page 110](http://html1.mh-extra.com/html/perkinelmer/aanalyst-200/aanalyst-200_user-manual_1530191110.webp)
Installation
5-22
Installing the Sample Tray
The sample tray is packed separately for shipping.
AAnalyst 200
Atomic Absorption Spectrome
ter
Figure 5- 11 Sample tray
1.
Hold the tray with both hands in the position shown in front of the instrument.
2.
Tilt the
tray mounting posts
downward at a slight angle as you insert them into the slots in the front
of the instrument.
3.
Slide the tray all the way into the front of the instrument base until it hits the stop. At that point,
gently push the tray edge downward to secure it into place. You will feel it lock into place.
NOTE: Do not block access to the E-Box located behind the right front cover. Removal of the power cord from
the appliance coupler will completely disconnect the AAnalyst 200 from AC power line.
This completes the procedure for installing the sample tray.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......