AAnalyst 200 User’s Guide
6-11
Installing Solvent-Resistant Components in the Burner
WARNING
Warning: Flammable Gases – Fire and Explosion Hazard
Not all gaskets in the burner unit are suitable for use with organic solvents.
Exposure to such solvents may cause the respective gaskets to swell. This may
result in a seepage of gas which is a serious fire hazard.
Before you aspirate organic solutions, remove the respective gaskets and install
solvent-resistant gaskets.
The standard o-ring in the burner end cap is not resistant to organic solvents. If you intend to aspirate
solutions containing organic solvents you must exchange this o-ring for a solvent-resistant o-ring in the
following way:
1.
Where applicable, extinguish the flame, allow the burner head to cool, shut down the gases at source,
and bleed the gas supply lines.
2.
Open the atomizer compartment door.
3.
Remove the burner head.
4.
If corrosive or toxic samples were run previously, pour a large quantity of deionized water slowly
through the neck of the burner chamber to thoroughly flush the inside and the drain system.
5.
Remove the nebulizer from the burner end.
6.
Remove the burner system from the spectrometer.
7.
Take the burner chamber apart.
8.
Remove the standard o-ring from the end cap and carefully fit the solvent resistant o-ring.
9.
Reassemble the burner system
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......