AAnalyst 200 User’s Guide
9-37
4.
Run one or more standards to determine if the capillary holder has been cleared. If it has
been cleared, proceed with normal spectrometer operation; otherwise, proceed with the next
procedure.
Complete Cleaning Procedure of AAnalyst Nebulizer
To clean a stubborn blockage from the nebulizer, perform the procedure below.
Removal and Disassembly the Nebulizer
1.
Turn off the spectrometer and turn off all gas supplies.
2.
Lift the end-cap spring clamp.
3.
Rotate counter clockwise and gently pull the nebulizer out of the burner end cap.
4.
After removing the nebulizer from the end cap,
carefully
remove the nebulizer gas supply
tubing (labelled NEB) from the nebulizer side arm as follows:
First loosen the connector nut (P/N N037-1485), see
WARNING
Wear protective gloves when you remove the nebulizer from the end
cap if you were analyzing toxic samples.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......