Cleaning the Burner System
The nature of the samples aspirated determines how frequently you need to clean the burner
system.
•
The burner should provide an even flame over the length of the burner slot. An uneven
flame may indicate that the slot needs cleaning. A badly clogged slot may cause a flashback
of the flame.
•
Aspirating samples with high solids content may cause deposits to build up along the burner
slot
.
•
Prolonged operation of the nitrous oxide-acetylene flame may cause deposits of pyrolytic
carbon to build up along the burner slot
.
WARNING
Unstable acetylides, which are likely to explode when dry, may be formed when
aspirating high concentrations of silver, copper, or mercury salts into an
acetylene flame.
WARNING
Warning: Hydrochloric and Nitric Acids – Risk of Burns
Hydrochloric (HCl) and nitric (HNO
3
) acids are corrosive and can cause
severe burns.
CAUTION
If high concentrations of Cu, Ag, or Hg have been aspirated into an acetylene
flame, clean the burner chamber immediately after use by following the
procedures in this section called Cleaning the Burner After Aspirating Cu,
Ag, or Hg sample.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......