3-9
Bench Requirement
The benchtop or area in which the AAnalyst 200 instrument is to be installed should be large
enough to accommodate the instrument and any accessories you’ll be using. The instrument itself
weighs 49 kg (108 lb.). If you are considering the use of other accessories, the bench you choose
to use should be able to sustain as much as 136 kg (300 lbs.).
Location
Typically, the system is installed on a laboratory work bench or table. The table should be
smooth, level, and free from vibration. The dimensions for the instrument are listed below and
shown in the next figure. For other accessories, refer to the manuals supplied with the accessory
.
The bench should be located near the required electricity and gas supplies. Allow additional
power outlets for all your accessories. A distance of at least 6 inches (153 mm) should be
surrounding both sides and rear of the instrument for proper ventilation. If not possible, the bench
on which the instrument is placed should be on wheels so that it can be easily moved.
Accessories
Make sure that there is sufficient room near the spectrometer for any accessories. For example,
we recommend that you place the (optional) flame autosampler and mercury hydride system on
the left side of the instrument or on a roll cart in front of the instrument.
Printer
Refer to the manuals supplied with your printer for dimensions. The printer may be placed on the
bench next to the instrument.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......