AAnalyst 200 User’s Guide
8-17
Table 8- 2 Suggested Standards
Element Standard
1
(mg/L)
Standard 2
(mg/L)
Standard 3
(mg/L)
Reslope
(mg/L)
Ag 0.2
0.5 1 0.5
Cu 0.5 1 1.5 1
Mg 0.05
0.1 0.2 0.1
Mn 0.2
0.4
0.6 0.4
Pb 2 5
10 5
3.
For samples 1, 2 and 3, use Standards 1, 2, and 3, respectively.
Set up the autosampler
The autosampler you intend to use must be connected to the spectrometer. Installation procedures
are described in the
User’s Guide
provided with the autosampler.
Set up the autosampler by following the directions given in Chapter 6
, Operation of the
Instrument.
Load the autosampler
In the autosampler tray, location zero (0) is reserved for the wash, (1) is for blank, and locations 2
through 9 are reserved for standard solutions. The last standard location is reserved for reslope.
If not already done, load the autosampler tray with a blank, standards, and samples.
Enter parameters
Refer to Chapter
Touch Screen
, section
Analyze Page: Autosampler
, if you require information
on how to enter analytical parameters. Fill in the entries for autosampler tray, first and last sample
location, and recalibration frequency.
If not already done, install and align the required lamp, and ignite the flame.
On Parameters page, Spectrometer subpage:
1.
Enter 1 second for the
Integration Time
.
2.
Enter 3 for the number of
Replicates
.
3.
Enter 3 seconds for the
Read Delay
.
4.
Enter the required
parameters.
Summary of Contents for aanalyst 200
Page 1: ...AAnalyst 200 User s Guide ...
Page 6: ......
Page 7: ...Introduction 1 ...
Page 8: ......
Page 18: ...Introduction 1 12 ...
Page 19: ...Safety Practices 2 ...
Page 20: ......
Page 44: ...Safety Practices 2 26 ...
Page 45: ...Preparing Your Lab 3 ...
Page 46: ......
Page 66: ...Preparing Your Lab 3 22 ...
Page 67: ...System Description 4 ...
Page 68: ......
Page 89: ...Installation 5 ...
Page 90: ......
Page 115: ...Operation of the Instrument 6 ...
Page 116: ......
Page 132: ...Operation of the Instrument 6 18 Figure 6 9 Installing the lamp compartment cover ...
Page 158: ...Operation of the Instrument 6 44 ...
Page 159: ...Touch Screens 7 ...
Page 160: ......
Page 173: ...AAnalyst 200 User s Guide 7 15 Screen 7 10 Spectrometer Subpage Flame Technique ...
Page 210: ...Touch Screens 7 52 ...
Page 211: ...Analyzing Samples 8 ...
Page 212: ......
Page 220: ...Analyzing Samples 8 10 Screen 8 4 Display Calibration screen ...
Page 231: ...AAnalyst 200 User s Guide 8 21 The results are displayed and printed Screen 8 12 Analyze page ...
Page 244: ...Analyzing Samples 8 34 ...
Page 245: ...Maintenance 9 ...
Page 246: ......
Page 268: ...Maintenance 9 24 Figure 9 6 Removing the burner assembly from the instrument ...
Page 291: ...AAnalyst 200 User s Guide 9 47 Figure 9 15 Removing the E Box from the instrument ...
Page 293: ...AAnalyst 200 User s Guide 9 49 Figure 9 16 Replacing the E Box ...
Page 307: ...Troubleshooting 10 ...
Page 308: ......
Page 320: ...Troubleshooting 10 14 ...
Page 321: ...Index ...
Page 322: ......