ALD
Ozone Delivery System
Oxford Instruments Plasma Technology
Equipment Manual
OPERATING
INSTRUCTIONS
Printed: 14-Jul-10, 7:13
Page 5-3 of 4
Issue 01: July 10
Turn the fast ALD valve (CH) on to admit ozone to the process chamber (see Figure 5-3). For
most processes, use a one second pulse of ozone.
Figure 5-3: Fast ALD valve is open
5.4
Faults
5.4.1
Fault displayed on the PC2000 page
Figure 5-4 shows the ozone panel if a fault has occurred.
Figure 5-4: A fault in the ozone delivery system
A red alert message is generated if the ozone generator is turned on but the generator does
not return an “OK” message within 30 seconds. The red alert stops the process that is
running. The alert message is “The ozone generator output has failed, please check the
generator, power connections and the ozone monitor”.
HAZARDOUS GAS
IF THE OZONE MONITOR DISPLAYS ELEVATED OZONE
CONCENTRATIONS, DO NOT OPEN THE CABINET. WAIT UNTIL A SAFE
LEVEL IS INDICATED ON THE OZONE MONITOR BEFORE OPENING THE
CABINET. IT CAN TAKE SEVERAL HOURS FOR THE OZONE
CONCENTRATION TO DECAY TO SAFE LAVELS.
If the ozone monitor displays an elevated ozone concentration, DO NOT attempt to open the
cabinet. Wait until the monitor indicates a safe ozone concentration before opening the
cabinet. Depending on the levels of ozone within the cabinet, this may take several hours.
If the ozone concentration has not decayed to a safe level after several hours, it is possible
that the ozone monitor has failed.
Perform the following steps to recover the situation:
1)
Check that the ozone generator is powered off.
2)
Put on appropriate personal protective equipment.
3)
Ensure that the area around the cabinet is well ventilated. If necessary, install
temporary extraction for the area.
4)
Open the cabinet door and fault-find the ozone monitor.