WARNING
BEFORE PROCEEDING WITH ANY MAINTENANCE WORK, READ SECTION 1 - HEALTH AND SAFETY.
Note that PFPE oil should never need changing if used as specified.
CAUTION
Changing from a PFPE oil to a mineral oil or vice versa should only be
attempted after reference to the manufacturers. If the oils are allowed to
come into contact with each other, they emulsify, with fatal results for the
pump.
6.7.2.3
Annually
Remove and clean the wire mesh filters (if fitted) on the intake ports, using a
perfluorinated liquid such as Freon when PFPE oil is in use, or Isopropyl alcohol (IPA)
when mineral oils (hydrocarbons) are in use.
CAUTION
When replacing filters, do NOT change filter type (filling material) without
reference to Oxford Instruments Plasma Technology.
Maintenance
Issue 4: July 10
Page 6-12 of 16