OIPT Systems
4.1
Mandatory Specifications for compressed air supplies
•
Inlet pressure to filter/mist
separator/regulator unit:
6 bar (90 psig) minimum.
OUTLET TO SYSTEM:
•
Oil content:
Less than 10 ppm
•
Maximum Moisture Content:
-3ºC (25ºF)
•
Filtration:
Maximum particle size of 0.3 microns
•
Regulator outlet pressure:
Adjustable from 3 bar to 6 bar (45 psig to 90
psig). 6 bar must be the maximum provided.
•
Maximum flow rate:
135 litres/minute (5 scfm).
•
Pressure monitoring:
0 to 10 bar (0 to 150 psi) pressure gauge.