System Manual
PKN
mçí=éêÉé~ê~íáçå=Ñçê=î~éçìê=Çê~ï=
1) Before connecting a source pot, ensure that the lines have been pumped.
2) Connect the pot in accordance with local health and safety measures and MSDS
3) Connect the inlet and outlet – DO NOT TURN THEM ON YET
4) Pump out atmosphere which is trapped during connection of inlet and outlet valves of the
source pot: (THE INLET VALVE OF THE SOURCE POT IS NOT CONNECTED.)
i)
The outlet is pumped out via the dose ALD valve D.
ii) Close D1, flow Argon from the purge MFC to clear any residual air in the delivery
line
iii) Repeat pump purge as required.
iv) OUTLET VALVE IS STILL SHUT
5) Degassing
Precursors are filled under inert gas at around 2 bar. This space has to be evacuated.
i)
Turn on D1 for 1 minute
ii) Gradually, open the outlet valve, note pressure rise does not trip the CM gauge.
iii) Shut manual valve
iv) Repeat steps ii) and iii) above until there is no pressure rise
v) Turn off D1
vi) Pot is now degassed
6) Start processing in vapour draw mode.
Precursor Source Pot
Issue 1: September 07
Page 12 of 20
Printed: 7-Oct-07, 12:13