![Oxford Instruments OpAL System Manual Download Page 115](http://html.mh-extra.com/html/oxford-instruments/opal/opal_system-manual_1664943115.webp)
System Manual
=
lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó=
lé^i
RKO
póëíÉã=éçïÉêJìé=
This sub-section describes the initial power-up (switch on) of the system. It assumes that the
system is switched OFF at the wall-mounted safety isolation box.
WARNING
BEFORE POWERING THE SYSTEM UP, CHECK THAT:
A)
THE ‘ON-SITE’ MAINTENANCE LOG SHOWS THAT THE SYSTEM IS IN A FIT STATE FOR
CUSTOMER OPERATION.
B)
ALL DOORS ARE CLOSED AND ALL COVERS ARE IN PLACE.
C)
THERE IS NO VISIBLE DAMAGE TO THE SYSTEM.
To power up the system, proceed as follows:
1)
Ensure that all manually operated cooling water taps are turned ON.
2)
Ensure that the compressed air supply is ON.
CAUTION
If the gas lines may contain air, do not open any gas taps. Continue with the start up
procedure, then use the system to evacuate the lines one at a time. When it is certain that the
air has been removed and there are no leaks, open the gas taps.
3)
If you are certain that the gas lines do not contain air, turn all the manually
operated gas taps (on gas cylinders etc.) to ON.
4)
Ensure that all covers and panels are fitted and secured.
5)
Open the console door, then set all Remote/Local switches on the panel-mounted
units (RF Generator, etc) to REMOTE.
6)
Switch ON all the panel-mounted units.
7)
Ensure that the serial communications cable from the remote PC system controller is
plugged into the ‘D’ socket located at the rear of the console.
8)
Check that the Emergency Stop button is OUT.
9)
Set the wall-mounted safety isolation box switch to ON (I).
10)
Ensure that the
Power On
LED (located on the system console) is illuminated.
11)
Operate
System ON
button (coloured green), located on the console front panel
behind the console door.
12)
Ensure that the
System On
LED (located on the system console) is illuminated.
13)
Switch the remote PC system controller ON.
Operating Instructions
Printed: 06 November 2007 09:42
Page 5-5 of 60
Generic (OpAL Plasma) Issue 1: Oct. 07
Summary of Contents for OpAL
Page 14: ......