System Manual
=
lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó=
lé^i
RKNMKV
j~ëë=Ñäçï=Å~äáÄê~íáçå=é~ÖÉ=
Fig 5-20: Mass flow calibration page
CAUTION
Some gas mixtures may produce particles in the chamber or gas
lines.
Check only ONE gas at a time.
This page is similar to the Leak Detection page (see sub-section 5.10.8, page 5-51) with the
addition of setpoint boxes for the Mass Flow Controllers. When the initial pumping and MFC
stabilisation period ends, the chamber seals and fills slowly. The rate-of-pressure rise is
calculated and displayed.
NOTE: Chamber pressure depends on quantity of gas added and on the chamber
temperature. If a high-power plasma has been run recently, the chamber will
be hotter and the rate-of-pressure rise will be greater for the same gas flow.
Operating Instructions
Printed: 06 November 2007 09:42
Page 5-53 of 60
Generic (OpAL Plasma) Issue 1: Oct. 07