System Manual
RKNMKNMKU
`Ü~ãÄÉê=îÉåíáåÖ=íáãÉë=é~åÉä=
Purge cycle
fields
The maximum length of time (hours, minutes and seconds) the
process chamber is purged before venting is continued.
Vent to atmosphere
fields
The maximum length of time (hours, minutes and seconds) the
process chamber is vented before a timeout condition is reached.
RKNMKNMKV
eÉ~íÉê=íçäÉê~åÅÉ=é~åÉä=
Tol%
– The percentage difference between the cryo temperature setpoint and readback
before an alarm condition is triggered.
Max Scale
– The maximum scale value (deg C).
RKNMKNMKNM
^m`=qçäÉê~åÅÉ=é~åÉä=
Pressure
fields
Tol%
– The percentage difference between the process chamber
pressure setpoint and readback before an alarm condition is
triggered.
Max
Scale
– The maximum scale value (Torr).
Stabilise Time
fields
The maximum length of time (hours, minutes and seconds) before
the process chamber pressure has stabilised.
Operating Instructions
Generic (OpAL Plasma) Issue 1: Oct. 07
Page 5-58 of 60
Printed: 06 November 2007 09:42