![Oxford Instruments OpAL System Manual Download Page 197](http://html.mh-extra.com/html/oxford-instruments/opal/opal_system-manual_1664943197.webp)
System Manual
lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó==
lé^i
WARNING
BEFORE PROCEEDING WITH ANY MAINTENANCE WORK, READ SECTION 1 - HEALTH AND SAFETY
.
TKRKO
d~ë=Äçñ=áåíÉêäçÅâë=
Refer to drawing: SE81B26657 (PC2003 gas pod loom).
To enable process gases, the RF interlock chain must be complete. The gas box interlock is
shown in Fig 7-2.
1
BLK
20
24VDC
LID
SWITCH
PCB 94-PC00S26866
REAR SERVICES
PANEL
SKT25W
PIN4
PL25W
PIN4
GAS BOX
SYSTEM CONSOLE
SERVICES
PANEL
SKT25W
PIN4
PL24W
PIN4
PCB 94-PC81S26655
JP6
PIN1
JP6
PIN2
JP6
PIN3
Via JP3 to CAN unit
controlling gases 1 - 4.
See NOTE.
NOTE: The function of the signal is to supply
24Vdc power to the digital outputs of
the CAN unit. The interlock continues
through idendical boards (where fitted)
which drive gases 5 - 8 and 9 - 12.
Fig 7-2: Gas box interlock chain
TKRKOKN
fåÅçãé~íáÄäÉ=Ö~ëÉë=
1
st
PCB Gas
2
nd
PCB Gas 3
rd
PCB Gas Type A
Type B
Type X
1
st
Gas
1 5 9 LK3A
+
4
LK3B
LK3A
2
nd
Gas
2
6
10
LK5A + 6
LK5B
LK5A
3
rd
Gas
3
7
11
LK7A + 8
LK7B
LK7A
4
th
Gas
4
8
12
LK9A + 10
LK9B
LK9A
Gases are designated as one of three types:
Gas type A: Typically oxidising gases (e.g. oxygen)
Gas type B: Typically fuel gases (e.g. hydrogen)
Gas type X: Gases normally miscible with most other gas types.
If ANY gas Type A is enabled, then ALL gas Type B lines are disabled.
The gas box has a facility to prevent incompatible gases from being enabled simultaneously,
using soldered links.
Troubleshooting
Printed: 25-Oct-07, 11:47
Page 7-11 of 12
Issue 1: August 07
Summary of Contents for OpAL
Page 14: ......