![background image](http://html1.mh-extra.com/html/omron/z500-series/z500-series_operation-manual_743394090.webp)
88
SECTI
ON 4
C
learing
Measurem
e
nt Item
s
Z500
Operation Manual
SECTION 4
Menus for Expert Menu
Step
5
Measurement Conditions
Set the measurement conditions. The following can be set in
Conditions
and its lower hierarchies.
Perform the setting as required.
Averaging number
Use this function to disregard sudden changes in the waveform. The average of
the setting number of times is output as the measurement result.
The asterisk (*) indicates the default setting.
• The Relationship between the Measurement Response Time and the Resolution
CHECK
The setting can be performed through the adjustment menu.
REFERENCE
Refer to page 28.
1.
Select
Conditions.
Menu
Description
Reference
# to avg
The average of the setting number of times is output as
the measurement result. Set this function to disregard
sudden changes in the waveform.
Page 88.
Trigger
From the measurement results that have been obtained
within a specified measurement period, the specified
characteristic quantity is held and output.
Page 90.
Details
Removing noises, outputting measurement failure, and
other miscellaneous methods can be set in more detail.
Page 96.
Selection
1, 2, 4, 8*, 16, 32, 64, 128, 256
Averaging
number
Short
1
256
Long
Low
High
Measurement
response time
Resolution
Image
Compensatn
Meas set
Conditions
Output
Scn 0 Set