System Settings
6-3 Environment Settings
Environment settings for the Z500's system can be made using the items shown below.
The asterisk (*) indicates the default setting.
Item
Selection
Description
Settings
Conversatn*
Expert
Select the menu used to set the measurement
conditions.
Sensor
Diffuse*
Mirror
Select the Sensor mounting method. Select the item that
corresponds to the mounting method used for the Sensor
connected. The default setting varies depending on the
Sensor model connected.
Scene
Console*
Terminals
Select whether scene changing is performed on the
Console or using the terminal block.
Monitor
Image*
Digital
Trend
Profile
Select the type of screen that is displayed when the
power supply is turned ON.
1st scene
Scn 0* to Scn 15
Select the scene number that is displayed when the
power supply is turned ON.
Dec. pts
1 to 5* digits
Select the lowest digit of measurement values displayed
on the monitor.
Download
ON
OFF*
The Z500 acquires information from the sensor at star-
tup. The startup time can be shortened by downloading
this information into the controller and storing it.
•ON
The information stored inside the controller is acquired
at startup. The startup time will be shortened by compar-
ison to when the OFF function is used.
•OFF
Information is acquired via the sensor connection at
startup. The startup time will be longer than when the
ON function is used.
NOTICE
To replace the sensor for connection to the controller,
perform the following procedure:
1. Turn OFF the download function.
2. Replace the sensor.
3. Download the new sensor information.
4. Turn ON the download function.
If the sensor is replaced when the download function is
ON, measurements cannot be taken correctly with the
new sensor, because the old sensor information (such as
the measurement range) is still stored.
Also, perform the above procedure when a sensor of the
same model is used to replace the existing sensor.