SECTION 4
Menus for Expert Menu
Step
6
Output Settings
The Z500 can transmit data to an external device in three formats; analog, terminal, and RS-232C. In
the example provided below, the procedure for setting the output contents to the external device is
explained. The output data and their transmission methods are as follows:
Analog
Measured values or profile data are converted to analog values and are output.
Setting items vary depending on the outputting data.
Outputting Measurement Values
Use the procedure explained below to set the measurement values correspond-
ing to the upper limit (20 mA/5 V) and lower limit (4 mA/-5 V) for analog output.
This setting enables output to be scaled (e.g., 1 mm corresponds to 1 mA) as
required. Set both the analog output and measurement values so that the upper
limit is greater than the lower limit.
The asterisk (*) indicates the default setting.
Data
Description
Analog
p.100
Terminals
p.103
RS-232C
p.105
Measure-
ment value
The measurement results for each output number are output
in numerical values.
supported
Not
supported
supported
Judgment
result
The judgment results for each output number are output.
Not
supported
supported
supported
Profile
Data on height of 126 lines are output.
When setting to
Analog
, analog waveforms are output.
supported
Not
supported
supported
Trend
Only the Trigger function allows this type of output.
Data that have been held for each output number are output.
Not
supported
Not
supported
supported
20mA
(+5V)
4mA
(-5V)
Measurement
lower limit
Measurement
upper limit
Measurement value
Analog
Item
Selection/Setting
Description
Output
OUT0 to OUT7
Profile
(OUT0*)
Selects the output number to be set.
4mA (-5V)
-9999.99999
to 9999.99999mm
Sets the lower limit for the analog output measurement range.
The default setting varies depending on the Sensor model connected.
20mA (+5V)
-9999.99999
to 9999.99999mm
Sets the upper limit for the analog output measurement range.
The default setting varies depending on the Sensor model connected.