Z500
Operation Manual
SECTION 3
Menus for Conversational Menu
When selecting
Fixed
1.
Select
Fixed
.
The screen for setting the sensitivity level will be displayed.
2.
Select the level using the
Up
and
Down
Keys while monitoring the image on the
monitor.
The level displayed first is the level
selected by the Z500 for the workpiece
being measured. Adjust the level while
monitoring the line brightness graph and
the received level on the monitor.
120 max.: Insufficient light density
255 min.: Excessive light density
3.
Press the
ENT
Key.
A confirmation message will be displayed.
4.
Select
Execute
when the basic settings
have completed.
Select
Settings
when continuing the
measurement contents settings.
Go to
Step
3
.
ENT:Set ESC:Back
Auto
Fixed
Multi
Scn 0 Set
Sen0's sensitivity method:
Select Fixed or Multi if
measurement unstable with Auto.
BRIGHT
DARK
FAR
NEAR
[ 15]
[ 170]
[ .]
[ ]
LV 15
ENT:Set ESC:Back
Sen0
LV
PEAK
LV
PEAK
Sen1
Sen0's sensitivity method:
Select Fixed or Multi if
measurement unstable with Auto.
Line Brightness
Received level
Settings will be registered.
Allocate measurement items
to OUT0 to OUT7
in the measurement settings.
Settings can be accessed
from the Adjustment Menu.
Execute
Cancel Settings