102
System Message
Cause
Recommendation
Catheter failure
Sudden pressure change that may occur during
implantation or removal of Flex Catheter.
Disconnect and re-connect Flex Catheter after completing
the implantation or removal process.
Catheter breaks during use.
Disconnect catheter and replace with new one (see
directions for use supplied with catheter).
Catheter fails during initialization.
Disconnect catheter and replace with new one (see
directions for use supplied with catheter).
Catheter is initializing
Camino ICP Monitor is in process of initializing
new catheter connection.
Wait until Camino ICP Monitor completes catheter
initialization process.
Cooling fan failure
Cooling fan inside Camino ICP Monitor has
stopped running.
Turn off Camino ICP Monitor to prevent overheating and
contact Natus Neuro.
Data export failed. Could
not detect or write to a
USB drive.
Problem with connection or compatibility of
USB drive.
No transfer of trend data via USB connection.
ICP above alarm limit
Patient’s Mean ICP value has exceeded user-
specified high ICP alarm limit for more than 5
seconds.
Respond to patient’s need immediately. Once the patient’s
Mean ICP value falls within the alarm limit, the alarm will
turn off.
ICP is out of accuracy
range
Mean ICP value has exceeded the range that the
Camino ICP Monitor can measure accurately
(between -10 mmHg to +125 mmHg).
If message appears during implantation or removal of the
Flex Catheter, disconnect and re-connect the Flex Catheter
to the Flex Extension Cable.
If message appears during patient treatment, do not rely on
ICP measurements until the Mean ICP value falls within the
accuracy range.
Monitor overheating,
check vent
Temperature inside Camino ICP Monitor has
exceeded 80 °C.
Check air vent on rear panel of Camino ICP Monitor. If
anything is blocking vent, remove it.
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...