128
Battery Charge
1.
Run the Camino ICP Monitor on battery power until the remaining battery charge is depleted. Note the
when the battery power falls to approximately 5 minutes or less of remaining charge, the monitor will
activate an audible alarm. After the alarm sounds for approximately 5 minutes, the monitor will
automatically shut down.
2.
Plug the monitor into an AC adapter outlet to start re-charging the battery.
3.
Press the
Power
button to turn on the monitor. On the touch screen, verify that the
battery power symbol that appears on the status bar displays the electrical charge
symbol.
4.
Press the power button again to turn off the monitor. Continue charging the battery to full capacity. The
time for re-charging the battery while the monitor is turned off will take less than 5 hours.
Notice
If the battery takes more than 5 hours to charge to full capacity, try replacing the battery with a new one. To order a
battery, contact Natus Neuro.
5.
Once the battery reaches full charge, the battery power symbol will display four green bars.
6.
Unplug the monitor and run it on battery power only and verify that the battery charge lasts
a minimum of 1 hour and 30 minutes. If the battery charge does not last a minimum of 1 hour and 30
minutes, contact Natus Neuro to order a new battery.
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...