117
1.
Plug the Camino ICP Monitor to an AC wall outlet.
2.
On the right side of the monitor, attach the Fiber Optic Catheter Cable and a functional Fiber Optic
Catheter to the
ICP (CAMCABL)
port.
3.
Turn on the monitor. On the touch screen, press the Main tab and verify the ICP value reads 0. If not,
use the black zero adjustment tool from the Fiber Optic Catheter kit to set the ICP value to 0 mmHg.
4.
Insert the Fiber Optic Catheter into the manifold.
5.
Connect the other end of the manifold to the pressure simulator.
6.
If necessary, adjust the pressure simulator to 0 mmHg.
7.
Verify that the ICP values on both the pressure simulator and the Natus monitor’s touch screen
read 0 mmHg.
8.
Adjust the pressure simulator control knob to the following pressures and verify the ICP monitor displays
the ICP value within the stated limits. It will also be necessary to take into account the settings accuracy of
the pressure simulator:
Applied Pressure (mmHg)
ICP Monitor (mmHg)
0
-2 to 2
20
18 to 22
40
38 to 42
60
56 to 64
100
94 to 106
125
117 to 133