Prior to implanting the catheter into the patient, disconnect the Fiber Optic Catheter from the Fiber Optic
Catheter Cable.
7.
After implanting the catheter into the patient, reconnect the catheter to the Fiber Optic Catheter Cable.
Warning
Once the Fiber Optic Catheter has been zeroed to the Camino ICP Monitor, do not replace the Fiber Optic Catheter Cable
being used for patient measurement. Replacing the Fiber Optic Catheter Cable with a different cable after the ICP catheter
has already been zeroed may result in inaccurate patient measurements.
Warning
Once the Fiber Optic Catheter has been zeroed to the Camino ICP Monitor, do not transfer this zeroed catheter to any
other monitor. Transferring a zeroed catheter to a different monitor may result in inaccurate ICP measurements.
Notice
If the patient only requires an ICP catheter, the temperature value on the touch screen will appear blank (with two
dashed lines).