109
Responding to System Failure Messages
If a system failure occurs, the Camino ICP Monitor will display a message window that identifies an error code and description of the problem. The
following table lists each error code associated with a system failure:
Error Codes
Cause
Recommendation
E0011, E2002, E2003, E2004, E2008, E2011, E2024, E2025,
E2026, E2032, E2037, E2041, E2042, E2044, E2045, E2046,
E2047, E2048, E2049, E2051, E2057, E2061, E2101, E2103, E2104,
E2105
Sensor board failure.
Not resolvable, contact Natus Neuro.
E0012, E1002, E1003, E1004, E1008, E1032, E1033, E1041, E1042,
E1044, E1045, E1046, E1047, E1050, E1057, E1061, E1101, E1103,
E1104, E1105
Power board failure.
Not resolvable, contact Natus Neuro.
E0013
General software error.
Not resolvable, contact Natus Neuro.
E0015
Incompatible firmware error.
Not resolvable, contact Natus Neuro.
E2051
Monitor powered on with USB drive
connected.
Remove the USB drive and restart the
monitor.