106
Problem
Cause
Recommendation
ICP waveform appears too small
or too large.
ICP waveform not scaled correctly.
On touch screen, press
Main
tab and use the
Scale
button to
adjust scaling factor.
ICP waveform appears as a solid
wave instead of a line wave.
Waveform Style is set to
Fill
mode.
On touch screen, press
Settings
tab and select
Set
Waveform->Line
.
ICP waveform appears as a line
wave instead of solid wave.
Waveform Style is set to
Line
mode.
On touch screen, press
Settings
tab and select
Set
Waveform->Fill
.
Time/date appears incorrectly.
Time and/or date is set incorrectly.
On touch screen, press
Settings
tab and select
Set Time
and Date
to specify current time and date.
Incorrect language is displayed.
Choice of language is set incorrectly.
On touch screen, press
Settings
tab and select
Set
Language
to specify language.
Discrepancy of ICP
measurements between the
Camino ICP Monitor and the
patient bedside monitor.
Patient bedside monitor is not synchronized to
the Camino ICP Monitor correctly.
On the Camino ICP Monitor, press
Synchronize to
Monitor
to re-calibrate the two monitors.
Incorrect patient bedside monitor adapter
cable being used.
Verify correct patient bedside monitor cable is being used.
PMIO cable or patient bedside monitor
adapter cable is faulty.
Replace PMIO cable or patient bedside monitor adapter cable.
PMIO cable or patient bedside monitor
adapter cable is loose.
Check cable connections between Camino ICP Monitor and
patient bedside monitor.
Patient bedside monitor has malfunctioned. Refer to manufacturer’s troubleshooting guide for the patient
bedside monitor.
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...