105
Problem
Cause
Recommendation
Touch screen freezes
System hardware error
•
Press and hold the power button for several seconds to
perform a forced shutdown of the system.
•
Press the power button to turn on the system.
Camino ICP Monitor does not
power on or off when power
button is pressed.
Firmware error
•
Disconnect AC adapter cord from Camino ICP Monitor and
remove battery.
•
Re-insert battery.
•
Reconnect AC adapter cord to Camino ICP Monitor.
•
Press the power button.
The Mean ICP value on the touch
screen displays two dashes (- -).
Poor connection between Fiber Optic Catheter
Cable and Camino ICP Monitor.
Make sure the cable is properly connected to the Camino ICP
Monitor.
Poor connection between Fiber Optic Catheter
Cable or Flex Extension Cable and the
catheter.
Replace the catheter (see directions for use supplied with
catheter).
Faulty catheter
Replace the catheter (see directions for use supplied with
catheter).
Damaged Fiber Optic Catheter Cable
Replace Fiber Optic Catheter Cable.
Damaged Camino Flex Catheter Cable.
Replace Camino Flex Catheter Cable.
The temperature value on the
touch screen displays two
dashes (- -).
Faulty catheter
Replace the catheter (see directions for use supplied with
catheter).
ICP only catheter in use
Replace the catheter with ICP/temperature catheter (see
directions for use supplied with catheter).
Damaged Fiber Optic Catheter Cable
Replace Fiber Optic Catheter Cable.
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...