Cleaning
Ensure that all the connections between the cables, catheter, cable strap, AC power adapter, USB-to-
RS232 serial adapter, and monitor have been removed before cleaning the Natus Camino ICP
Monitor.
Limits on
Reprocessing
The Fiber Optic Catheter Cable may be sterilized thirty (30) times and cleaned/ disinfected one
hundred (100) times. Do not use the cable after reaching these reprocessing limits. For tracking
purposes, use the reprocessing record sheet (doc # 025471) supplied with your Fiber Optic Catheter
Cable.
The Flex Extension Cable may be sterilized thirty (30) times. Do not use the cable after reaching this
sterilization limit. For tracking purposes, use the reprocessing record sheet (doc # 026426) supplied
with your FLEX Extension Cable.
There are no limits on reprocessing the Natus Camino ICP Monitor and cable strap.
Recommended Manual
Cleaning Method (not a
sterilization procedure)
Using either 70% IPA with a lint free wipe or a Super Sani-Cloth® (or equivalent solution), thoroughly
wipe all surfaces at least three (3) times and then inspect the surfaces for visible residues. If residues
remain, use a new Super Sani-Cloth (or equivalent solution) or lint free wipe soaked with 70% IPA and
continue wiping the surfaces until they are visibly free of residues. If the Flex Extension Cable or the
Fiber Optic Catheter Cable cannot be thoroughly cleaned, discard and replace the cables.
Recommended
Disinfection Method
Using either 70% IPA with a lint free wipe OR Super Sani-Cloth (or equivalent solution), thoroughly
wipe all surfaces once. Ensure that the surfaces remain visibly wet for a minimum of two (2) minutes
and use additional wipe(s), if needed, to ensure continuous two (2) minute wet contact time. Let the
Natus Camino ICP Monitor and any of its system components air dry prior to reuse.
Inspection and
Functional Testing
After each reprocessing event, visually inspect the Natus Camino ICP Monitor and any of its system
components for any wear and tear.
Containment and
Transportation
It is recommended that the Natus Camino ICP Monitor and any of its system components be cleaned
and disinfected as soon as is reasonably practical after use.
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...