134
Item
Specification
Operating Pressure
700 to 1060 hPa
Shipping / Storage Pressure
500 to 1060 hPa
Operation Limits
•
Temperature = 15 °C to 40 °C
•
Humidity = 30% to 75% relative humidity, non-condensing
Shipping / Storage Limits
•
Temperature = -20 °C to 50 °C
•
Humidity = Relative humidity ranging from 25% to 80%.
Accuracy in Long Term
Operation
ICP catheter linearity with Camino Flex Catheter:
•
Range -10 to 10 mmHg: ±2 mmHg
•
Range 11 to 33 mmHg: ±3 mmHg
•
Range 34 to 125 mmHg: ± (6% + 1 mmHg)
ICP catheter linearity with fiber optic catheter:
•
Range -10 to 50 mmHg: ±2 mmHg
•
Range 51 to 125 mmHg: ±6%
Temperature catheter: ±0.3 °C
Rated ICP Accuracy Range
-10 to 125 mmHg
Rated Temperature Accuracy Range
(Direct Mode measurement)
30 °C to 42 °C
Protection Against Electric
Shock
Class II, type CF, defibrillation proof applied parts: ICP catheter connections, temperature
catheter connections.
Protection Against Harmful
Ingress of Water
IPX0
Mode of Operation
Continuous
HF surgical equipment
In compliance with the requirements of IEC 60601-2-49
Defibrillator recovery
0-15 Sec
Fire Hazard
Not suitable for use in the presence of flammable anesthetics mixture with air, oxygen, or nitrous
oxide.
Languages
English, Danish, Dutch, French, German, Italian, Polish, Spanish, Portuguese (Brazilian), Chinese (Simplified),
Japanese, Korean, Russian
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...